Product manual
Product use, preparation, operating conditions and storage.
Sequential coupling through the dye’s free amine for europium-cryptate labels used in time-resolved fluorescence.
Packs & quoteNH2 is a primary amine on the dye, not an NHS ester. Two unactivated primary amines do not couple directly. Use carboxyl activation, or react the dye amine with the NHS end of SMCC before coupling its maleimide end to a target thiol.
Use a TRF instrument with suitable excitation, delay and integration windows. Compare buffer blank, unlabeled sample and labeled product at common settings; ordinary steady-state intensity does not replace delayed detection.
Match the acceptor, proximity and detection windows when building a donor–acceptor system, with donor-only, acceptor-only and nonbinding controls. Luminescence from a Eu label alone does not establish energy transfer.
Use parameters for the exact catalog number when preparing reagents, setting reagent input or calculating results. Request a lot-specific CoA for test results.
The dye carries NH2; the target needs an activated carboxyl or a group matched to the linker. Do not add it to a native antibody as though it were an NHS dye.
Define the buffer and separation method for each step to limit linker reactions with unintended sample components.
Confirm the required free thiols and optimize reduction on a small sample if needed. Remove free reductant and proceed promptly to limit reoxidation.
React the Eu dye amine with the NHS end of SMCC or a suitable linker, then remove free linker. Select separation for the Eu intermediate’s molecular size and monitor recovery; protein ultrafiltration or desalting conditions do not automatically retain it.
React the purified maleimide intermediate with target thiols. Select pH, time and input for target integrity and linker requirements, then remove free Eu, linker and any blocking reagent.
Confirm the conjugate, sample recovery and binding, comparing label and controls at common delay and integration settings. Labeling calculations require correction and extinction data for this catalog number and the actual protein.
Select a document and enter your phone and email. Include the lot number for a CoA.
Product use, preparation, operating conditions and storage.
Spectra, measurement conditions and detection settings for the exact catalog number.
Lot-specific test results. Include the lot number from the vial.
Handling, storage and emergency response information.
Omit activator or linker, purify identically and assess adsorption background.
Use an unlabeled aliquot from the same preparation to assess changes in function, aggregation and recovery.
Include free-dye or no-target-handle controls to distinguish conjugation from adsorption. Match exposure, gain or voltage during detection.
Evaluate signal and background at matched delay and integration settings, recording instrument mode, excitation band, emission window and reading time. Recheck working concentration and background after changing the labeling form.
Include your sample, formulation and detection method when requesting application guidance.
Thermo Fisher Scientific · SMCC and Sulfo-SMCC · MAN0011295 Rev. C.0
Distinguishes the partners and reaction order of the NHS and maleimide ends and removal of free linker, supporting bridging routes for NH2 dyes.
Shafer DE, Inman JK, Lees A. · Analytical Biochemistry · 282(1):161–164 · 2000
Examines TCEP interactions with reactive groups, informing evaluation of reductant carryover and buffer exchange.
Mathis G. · Clinical Chemistry · 39(9):1953–1959 · 1993
Describes europium-cryptate donors, energy transfer to an acceptor and time-resolved signal correction, informing donor, acceptor and background controls.
Revvity · HTRF Principle
Explains Eu donors, compatible acceptors, proximity-dependent transfer and delayed detection. A Eu dye alone is not a complete TR-FRET assay; HTRF kit settings are not interchangeable product specifications.
No. NH2 supplies a free primary amine; NHS supplies an activated ester that reacts with an amine. Their partners and workflows differ.
Two ordinary primary amines do not couple directly. Use an appropriate activation or bridging route.
This depends on the reaction design. Residual linker can consume target sites or cause self-crosslinking; validate before simplifying the workflow.
TR-FRET requires a matched donor–acceptor pair, delay and dual-channel detection. Check instrument mode and controls rather than inferring FRET from Eu single-channel intensity.